73 Series Spectrophotometer
Accessory fitting and operation manual
061 672
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - 73 Series Spectrophotometer

73 Series SpectrophotometerAccessory fitting and operation manual061 672

Page 2 - Safety

10Confirmation will be needed to start the sipper pump. Press the key adjacent to the tick icon to confirm and start the sipper pump. Press t

Page 3

11To fine tune the amount of sample uptake press the key below the plus or minus icon to increase or decrease the amount of sample taken up. The recor

Page 4

13To perform a measurement place the sipper tubing into the sample and press the key below the sipper pump icon.Confirmation will be needed to start

Page 5

The pump direction is displayed by an arrow icon below the sipper peltier icon. The combined sipper peltier pump combines the functionality of t

Page 6 - (pumping)

Bibby Scientific Middle East LtdBPO Box 27842Engomi 2433Nicosia CyprusTel: +357 22 660 423Fax: +357 22 660 424e-mail: sales@bibby-scientificme.comwww.

Page 7

2 SafetyPlease read this information carefully prior to installing or using this equipment.1. The unit described in this manual is designed be operat

Page 8 - 0.000 ABS

3 SECTION 1 – Accessories and Spare Parts1.1 OPTIONAL ACCESSORIESPart Code Description of Accessory660 101 Internal printer735 401 Automatic 8

Page 9 - -> 120.5

4Squeeze the grey plastic clips together so that the printer top opens. Slot the printer into the top of the instrument and push down until it fits flus

Page 10

1.2.3.1 Automatic 8 cell turretTake the 8 cell turret base plate. Connect the power supply in the bottom of the sample chamber to t

Page 11

61.2.3.3 Sipper pumpFor this accessory as well as removing the passive accessory base plate, the front panel of the instrument must also be r

Page 12

77. Cut a small length of the sipper pump tube and push this over one end of the capillary tube. Connect this to the inlet port of the flow-through cuv

Page 13

81.3 USING THE ACCESSORIES1.3.1 Automatic 8 cell turretWhen the automatic 8 cell turret is in use the 8 cell turret icon is displayed in the bottom

Page 14

1.3.2 Peltier When the peltier is in use the peltier icon is displayed in the bottom right hand corner of the screen. The current temperature is dis